Chlorophyll Fluorescence

The capacity of a plant to carry out photochemistry is limited and will depend upon a range of factors including stresses caused by environmental conditions. Absorbed light energy in excess of that used for photochemistry must be effectively dissipated by non-photochemical processes. Such processes include the emission of heat and re-emission of small but diagnostically significant amounts of the absorbed radiation as longer wavelength red/far-red light energy. This re-emission of light is termed chlorophyll fluorescence.

Although chlorophyll fluorescence emission from whole leaf systems is too weak to be viewed with the naked eye, it can be observed from illuminated extracts of a chlorophyll solution. Peak chlorophyll fluorescence occurs in the red region of the spectrum (685 nm) and extends into the infra-red region to around 800 nm.

Each of these processes operate in direct competition for a finite pool of absorbed energy, any change in energy utilisation by one process produces a complementary change in the others. This fact enables chlorophyll fluorescence to be used as a rapid and reliable non-invasive probe of photochemistry.

Specialist equipment is required for the analysis of the chlorophyll fluorescence signature. A Chlorophyll Fluorimeter (or fluorometer) is designed specifically to detect the chlorophyll fluorescence emission from a sample. There are several different types of chlorophyll fluorimeter available.

Hansatech Instruments design and manufacture chlorophyll fluorimeters based around 2 different measurement techniques:

  • Continuous Excitation or Prompt Chlorophyll Fluorimeters
    A continuous excitation fluorimeter is designed to measure the Kautsky Induction or Fast Chlorophyll Fluorescence Induction (Kautsky and Hirsch, 1931*). The systems use focused, high intensity light from red LED’s to induce a fast chlorophyll fluorescence response from a dark adapted sample. Continuous systems require the use of a special leafclip system. This is a multi-purpose tool which provides dark adaptation for the sample (required for the measurement of maximum photochemical efficiency), defines the measurement area on the sample and prevents ambient light leakage into the highly sensitive photodiode used by the instrument for chlorophyll fluorescence detection.

  • Pulse-Modulated Chlorophyll Fluorometers
    A modulated chlorophyll fluorometer uses sophisticated electronics to separate chlorophyll fluorescence from ambient light. The systems achieve this using a rapid pulsing excitation light in order to induce a corresponding pulsed fluorescence emission. The fluorometer uses a highly sensitive photodiode to detect and record the pulsed fluorescence signal and to ignore any non-pulsed signal